The Nortec 2000D+ offers a frequency range of 50 Hz to 12 MHz for applications ranging from detection of cracks in tubes or structures to the discovery of minute flaws in aircraft materials. Single or dual frequency operation, easy to use digital conductivity, and rotating scanner support make the flaw detector ideal for numerous aerospace NDT applications.
Customer-interchangeable displays offer excellent visibility in any lighting condition. A VGA output drives a heads up display for inspection where conditions may be cramped or a large desktop monitor or projector for classroom training.
At less than 4 lbs (1.8 Kg), the rugged Foreign Object Debris (FOD) free case will survive harsh field and production environments. An adjustable tilt bail and anti-slip bumper allow the unit to be placed on just about any surface.
The Nortec 2000D+ incorporates our unique PowerLink™ software, which provides automatic probe recognition and documentation. The instrument can be set-up by recalling the program stored in the PowerLink™ chip, providing integrity and repeatability of inspection results.
As many as 120 programs can be stored and recalled later. Date and time are recorded with each set-up and are easily identified with alphanumeric values up to 29 characters long. 20 memory locations are available to store eddy current displays..
• 50 Hz - 12 MHz frequency range
• Single Li-Ion battery
• Lightweight, less than 4lbs. (1.8 Kg)
• Digital conductivity in International Annealed Copper Standard (%IACS) or Mega Siemens per meter (MS/m)
• Non-conductive coating readings in inches or millimeters
• Multiple scanner support
• Internal balance loads for single coil probe support
• Customer interchangeable displays:
- Hi-Brite Electroluminescent
- Monochrome Liquid Crystal
- Color Liquid Crystal
• Dual Frequency
• VGA Output
• Display Freeze to hold flaw signals
• PowerLinkTM Technology - automatic probe recognition and instrument set-up
• Split screen presentation with color coded soft keys
• On screen reference memory for go/no go applications
• On board storage of 120 programs
• Waterfall display
• Windows-based EddyMaster™ Software
BASIC PERFORMANCE |
FrequencyRange: 50 Hz - 12 MHz |
Gain: 0 - 90 dB in 0.1 dB steps. The horizontal and vertical gains may be adjusted separay or together. |
Rotation: Variable 0º - 359º |
Sweep: Variable from 0.005 - 4 seconds per division |
LowPassFilter: 10 - 500 Hz and wide band |
HighPassFilter: Off, 2 to 500 Hz. 2 pole response |
Probe Drive: 2, 6, 12 volts |
Variable Persistence: Screen persistence can be varied from 0.1 - 5 seconds. |
Probe Types: Absolute and differential in either bridge or reflection configuration. The instrument is fully compatible with NORTEC PowerLink™ probes. |
Alarms: Can be set to trigger on positive or negative |
Alarm Modes: 1-3 box gates, polar, sweep, conductivity, and coating thickness |
Trace Storage: 20 traces can be stored for recall. The traces can be static or frozen. They can contain up to 60 seconds of movement. The traces are stored with the date and time of capture. |
Program Storage: 120 instrument set-ups may be stored and recalled. The date and time of storage is recorded with each set-up. |
Print Out: Provides a custom configurable report header containing the display screen data and probe parameters including serial number (PowerLink™ probes only) |
Printers: Any serial printer |
Inputs / Outputs |
Power: 7-pin connector to charge the internal battery and operate the instrument on AC power |
RS-232: DB-9P connector, bi-directional serial data via RS-232 |
Probe Connector: 16-pin LEMO |
Analog Outputs: Horizontal and Vertical outputs of both F1 and F2. +/- 5 volts, 1 volt per division |
Alarm Outputs: 9-pin analog and alarm output connector |
VGA Output |
General |
Dimensions: 9.5" L x 5.5" H x 3.6" D (215 mm x 165 mm x 92 mm) |
Weight: 3.8 lbs. (1.7 Kg) with battery |
Display: Customer-interchangeable QVGA displays (320 x 240 pixels): color or monochrome LCD, Hi-Brite electroluminescent |
Operating Temperature: 14° F to 122° F (-10° C to 50° C), depending on configuration |
Storage Temperature: -60° F to 160° F (-51° C to 71° C), depending on configuration |
Humidity: 5 to 95% |
Classification: Based on Class 2 specifications from the MIL-PRF-28800F handbook |
Altitude: Maximum operating and non-operating altitude - 15,000 ft. (4600 m) |
Hazardous Area Operation: Safe operation as defined by Class I, Division 2, Group D, as found in the National Fire Association Code (NFPA 70), Section 500, and tested using MIL-STD-810F, Method 511.4, Procedure 1
Power
Power Requirements: 85 to 240 volts, 50-60 Hz. External holder charges batteries outside the instrument. Charging time is typically 4 hours.
Low Battery Protection: Display bar graph "gas gauge" indicates approximate operating time.
Battery Operating Time: 8 hours (nominal depending on configuration)
Conductivity
Frequency: 60 kHz or 480 kHz
Probe Type: NORTEC Conductivity Probe
Digital Conductivity Specification: Digital conductivity display from 0.9% to 110% IACS or 0.5 to 64 MS/m. Accuracy within +/- 0.5% IACS from 0.9% to 65% IACS and within +/- 1.0% of values over 62%. Meets or exceeds BAC 5651 specifications.
Non-Conductive Coating Thickness: Can measure non-conductive coating thickness from 0" to 0.015" (0 to 0.38 mm). Accuracy of +/- 0.001" (0.025 mm) over 0.00 to 0.015" (0 to 0.38 mm) range
Scanners
Scanner Compatibility: Will operate all Olympus Nortec and other commercially available scanners
Waterfall Display: Stores up to 60 sweeps per hole and includes an on screen readout of the distance to the defect from the start of the scan (PS-5 only)
Dual frequency
Frequency Extension: 50 Hz to 12 MHz
Second Frequency: 50 Hz to 3 MHz, 2nd frequency is an exact division of the first frequency in ratios of: 1/2 (F1 < 6 MHz), 1/4, and even divisors to 1/32.
Display: Frequency 1 (F1) only, Frequency 2 (F2) only, sum of F1 and F2, difference between F1 and F2, split screen with selected combinations of F1 and F2 and mixed frequencies
High Pass Filter: Available in Frequency 1 (F1) only
Standard Inclusions
Operating Manual
Power Cord
Li-Ion Battery
Universal Barrery Charger
Carry Case
Protective Soft Cover
VGA Cable
Calibration Certificate
*All specifications are subject to change without notice.
請(qǐng)輸入賬號(hào)
請(qǐng)輸入密碼
請(qǐng)輸驗(yàn)證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性,、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任,。
溫馨提示:為規(guī)避購(gòu)買(mǎi)風(fēng)險(xiǎn),,建議您在購(gòu)買(mǎi)產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。