目錄:上海鉑勒機電設(shè)備有限公司>>水質(zhì)分析儀器>>其他水質(zhì)分析儀及配件>> 6200010
The APA 6000 High Range Hardness Analyzer measures high-range hardness using USEPA-approved calmagite chemistry. Range 10 to 1,000 mg/L. Accuray and continuously measures up to two sample streams. Operates unattended for one month. Makes your water softening system more efficient and less costly.
Using a precise colorimetric method, the instrument performs an analysis every four minutes. It delivers continuous, accurate, and dependable performance with minimal cost for operation and maintenance.
Two separate sample streams can be monitored with the APA 6000 Hardness Analyzer when used with additional sample sequencing kits.
The interface of the APA 6000 Hardness Analyzer is menu-driven with on-screen prompting, self-test diagnostics, security provisions, and flexible data logging options. Simple menus (available in English, French, German or Spanish) guide the user through each step of operation.
The APA 6000 Hardness Analyzer can be used to monitor both influent and effluent in industrial applications and water treatment processes including demineralizer effluent, boiler feed-water, boiler water, and process water.
The APA 6000 High-Range Hardness Analyzer is designed specifically for reliable and continuous analysis at high concentrations. Use it to establish an automatic system for monitoring softener performance in industrial, ultrapure, and water treatment processes.
APA 6000 HR Hardness analyzer, installation kit, one month's supply of reagents, a maintenance kit, a sample conditioning kit, Hach user manual.
Accuracy:± 5 % of reading or ± 2 mg/L CaCO3 whichever is greater
Alarm:Two SPDT relays with contacts rated for 5A resistive load at 230 Vac. Additional relays available with optional Signal Output Modules.
Certifications:NRTL certified to UL and CSA standards and CE approved
Cycle Time:8.2 min
Drain Fitting:3/4 NPT barbed hose fitting
Inlet Pressure:0.5 to 30.0 psig (0.03 to 2.04 bar)
Length:21 in ( 533 mm ) (522 mm)
Outputs:Two 4-20 mA outputs suitable for recorders or PID control. Ouput span programmable over any portion of the measuring range (130 Vac isolation from earth ground)
Power Requirements (Hz):50/60 Hz
Power Requirements (Voltage):90 - 240 V AC
Range:10 to 1000 mg/L as CaCO3
Ranges:to as CaCO3
Repeatability:± 5 % of reading or ± 2 mg/L CaCO3, whichever is greater
Response Time:Less than 17 min for 90% response
Sample Flow Rate:max 100 to 2000 mL/min
Sample Temperature:5 to 50 °C (41 to 122°F)
Weight:56.22 lbs. (25.5 kg)
(空格分隔,最多3個,單個標(biāo)簽最多10個字符)