目錄:上海鉑勒機電設備有限公司>>水質(zhì)分析儀器>>其他水質(zhì)分析儀及配件>> D3455A6
Nominal Cell Constant (K) = 0.05, Sanitary (Cip) 1-1/2 Inch Flange, 6M (20 ft) Analog Cable, 316 Stainless Steel Electrode, Includes Digital Gateway And 1M (3.3 ft) Digital Extension Cable
These enhanced performance sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Simply key in this constant (Hach's easy DRY-CAL™ method) when configuring the analyzer to ensure the highest possible measuring accuracy. Also, each sensor has a Pt 1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1°C accuracy.
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach's sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
Compression fitting sensors Feature titanium electrodes and a compression fitting for universal installation with up to 4 inches (102 mm) insertion depth. The 1/2-inch or 3/4-inch male NPT compression fittings are offered in Kynar® (PVDF) or 316 stainless steel.
There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
Includes: sensor with cable, digital gateway, extension cable and manual
Accuracy:± 2 % of reading above 200 µS/cm
Measuring Range:to
Operating Temperature Range:-20 to 200 °C
Pressure Limit:0 to 300 psi (20.7 BAR)
Repeatability:± 0.5 % of reading
Response Time T90:30 s of step change
Transmission Distance:100 m